SKC

Parallel Particle Impactors (PPIs)

The parallel particle impactor (PPI) is a small sampler with 4 impactors in 1 unit, which provides the closest match to the ACGIH/ISO/CEN respirable and thoracic conventions.

Ask an engineer

These Parallel Particle Impactors (PPIs) are high performance, reusable thoracic and respirable PM samplers. The patented design [*]{US Patent No. 7,073,402} has the power of 4 impactors in one small, conductive aluminium sampler and weighs only 93g. The PPI belongs to the particle size selective sampling range.

How does a parallel particle impactor work?

The four small impactors sit in the inlet section. Each has a unique 50% cut-point to target one-quarter of the ISO/CEN curve, giving a precise fit along the entire curve. A sample pump pulls air through the inlet nozzle of each impactor. Particles larger than that cut-point are retained on the porous oiled impaction substrate, while smaller particles continue to the standard 37 mm collection filter for analysis. Disposable pre-oiled porous plastic substrates trap larger particles firmly, which reduces the effects of particle buildup and bounce. Unlike other samplers, the sampling efficiency of the PPI is not dependent on the collected particle type.

Which PPI sampler for which flow rate?

  • 8 L/min respirable (225-383): enhanced sensitivity for WEL/TWA silica and other compounds with low target concentrations, ideal for short term task monitoring; referenced in MDHS 14/4
  • 4 L/min respirable (225-382): higher flow option and use of intrinsically safe personal pumps
  • 2 L/min respirable (225-380): ideal for standard WEL/TWA sampling
  • 2 L/min thoracic (225-381): ideal for metal working fluids, or WEL/TWA monitoring for sulphuric acid, or other compounds with a thoracic limit; referenced in MDHS 14/4

Which filters and substrates does the PPI need?

It takes a 37 mm Quartz filter (5.0µm pore size), a 37 mm PTFE Filter (2.0 µm pore size, NIOSH 5524) or a 37 mm MCE Filter (0.8 µm pore size), on a cellulose pad or stainless steel screen for support. Four pre-oiled impaction substrates are required for each sample. Plastic disposable Parallel Particle Impactors (PPIs) are also available.

* US Patent No. 7,073,402