HAZ-DUST I Particulate Monitor
HAZ-DUST I particulate monitor: a portable survey tool that gives instantaneous readings for respirable dust on two selectable measuring ranges.
The HAZ-DUST I particulate monitor is a portable dust survey tool with simple operation and an instantaneous, easy-to-read display. It is highly sensitive yet lightweight and rugged. It is one of our particulate monitors.
What is the HAZ-DUST I used for?
Use it to identify areas with high levels of respirable dust and to evaluate dust generation and suppression. Readings are instantaneous and shown in mg/m3. Two measuring ranges are selectable: 0.01 to 200 mg/m3 and 0 to 20 mg/m3. The particle size range is 0.1 to 50 µm.
How does the HAZ-DUST I measure dust?
It uses near-forward light scattering. Airborne particles entering the instrument pass through an infra-red light beam and scatter the light. The amount of light that reaches the detector is directly proportional to the aerosol concentration.
What does the HAZ-DUST I offer for logging and unattended use?
The monitor operates for over 8 hours on its rechargeable battery and has a built-in battery test facility. Options:
- a data logger, supplied with DustComm Pro Software for data downloading, trend analysis and reporting of concentration changes over time;
- a tripod with adjustable mounting for unattended monitoring.
If you need a gravimetric pump together with the optical sensor, see the HAZ-DUST particulate monitor with gravimetric pump.
| Parameter | Value |
|---|---|
| Accuracy | ±10% to NIOSH 0600 |
| Sensing Range | 0.01 to 200 mg/m3, 0 to 20 mg/m3 |
| Display Resolution | 0.01 mg/m3, 0.1 mg/m3 |
| Particle Size Range | 0.1 to 50 µm |
| Calibration | NIOSH 0600 gravimetric reference – NIST traceable SAE fine test dust |
| Displayed Particle Concentration | Instantaneous |
| Power | Rechargeable NIMH battery |
| Operating Time | ≥8 hours |
| Charging Time | 10 to 12 hours |
| Temperature | Operating: -20 to +60°C Storage: -20 to +60°C |
| Dimensions | 23 x 7.6 x 3.8 cm |
| Weight | 600 g |
| Description | Part Number | |
|---|---|---|
| HAZ-DUST I Monitor Kit, includes: Monitor, 110-240 V AC Charger and carry case (requires calibration with equipment, sold separately) | 770-1100 |
| Description | Part Number | ||
|---|---|---|---|
| Calibration Standard, for verifying span and optical sensor performance. A light-scattering device providing a constant value. | 770-102 | | |
| Zeroing Accessory, for pumping filtered air into the sensor for a zero reading in contaminated environments | 770-130 | | |
| Data Logger: data storage, on-screen display of statistics and user-selectable alarm threshold, includes analogue signal cable, computer RS-232 interface cable and DustComm Pro Software on a USB Memory Stick. Dimensions: 17.8 x 8.8 x 6.3 cm (7 x 3.5 x 2.5 inches), Weight: 450 g | 770-113 | | |
| Analogue Signal Cable for Data Logger | 770-107 | | |
| Replacement NiMH Battery Pack for Data Logger | 770-127 | | |
| Tripod Stand, adjustable stand (1 to 1.5 m) and mounting plate for unattended monitoring | 770-129 | | |
| Cleaning Kit, includes bulb, swabs and lens tissue | 770-305B | ||
| AC Charger 110-240 V, can also be used to operate HAZ-DUST I for extended running time | 770-131 | ||
| Replacement NiMH Battery Pack for HAZ-DUST I | 770-108 | |
Questions about this product?