SKC

Anasorb Sorbent Tubes

SKC Anasorb sorbent tubes: the Anasorb name is used for SKC proprietary sorbents of all types, including Anasorb 747, CSC and GCB.

Ask an engineer

SKC Anasorb sorbent tubes are filled with Anasorb® sorbents; the Anasorb® name is used for SKC proprietary sorbents of all types. They are part of the range of SKC sorbent tubes.

To select a tube for a specific compound refer to the SKC Sampling Guides or use our SKC online Hazard Database

Which Anasorb sorbent samples polar compounds such as ketones?

Anasorb® 747 is a synthetic carbon with low ash content which studies showed as a candidate material for sampling polar organic compounds that cannot be sampled using coconut charcoal, such as ketones. Anasorb 747® has properties similar to the best petroleum charcoals and can be used for a very wide range of polar and non-polar organic vapours. When combined with a solvent appropriate to mixtures such as dimethylformamide, successful sampling is possible for a majority of organic vapours. Part number 226-81A is an Anasorb 747 tube.

How do the other Anasorb grades compare with Chromosorb, Carbotrap and Carbopack?

  • Anasorb® 727 is a microporous polymer for sampling less volatile reactive and polar compounds. It is derived from cross-linked styrene and similar monomers, and its sorbent properties are similar to Chromosorb® 106. The surface area is large, very unreactive, and extremely hydrophobic. See the Chromosorb sorbent tubes.
  • Anasorb® 708 is equivalent to Chromosorb® 108.
  • Anasorb® C300 sorbent is equivalent to Hydrar, Carulite.
  • Anasorb® CSC sorbent has a high surface area for sampling a broad range of primarily non-polar compounds; part numbers 226-09 and 226-73 are Anasorb CSC tubes. See the charcoal sorbent tubes.
  • Anasorb® GCB1 – Graphitised Carbon Black for sampling aliphatic and aromatic hydrocarbons and is equivalent to Carbotrap® B (20/40) and Carbopack B (60/80). This sorbent has proven to be valuable for sampling compounds of intermediate to high volatility.
  • Anasorb® GCB2 – Graphitised Carbon Black for sampling less volatile compounds. This sorbent has a low surface area and is equivalent to Carbotrap® C (20/40) and Carbopack C (60/80), and is useful in collecting semi-volatile organic compounds (SVOCs).

What is the difference between purged and unpurged tubes?

Note that thermally purged sorbent tubes are for immediate use in the field. Unpurged tubes are purged by a laboratory as needed.

Questions about this product?

Our application engineers aim to reply within 3 business hours.

Ask an engineer