Sioutas Personal Cascade Impactor

The Sioutas Personal Cascade Impactor separates and collects airborne particles in five size ranges, highly efficiently and without impaction grease or substrate coatings.

Ask an engineer

The Sioutas Personal Cascade Impactor (100-3002) separates and collects airborne particles in five size ranges, from coarse (>2.5µm) to ultrafine below 0.25 µm. It samples coarse, fine and ultrafine particles simultaneously. With PTFE filters it is highly efficient at collecting particles without impaction grease or substrate coatings, and at retaining unstable compounds for size-fractionated chemical analysis.

How does a Sioutas cascade impactor separate particle sizes?

The cut-points are 2.5 µm, 1.0 µm, 0.5 µm and 0.25 µm. Particles above each cut-point collect on a 25 mm filter in the matching stage; particles below 0.25 µm collect on the optional 37 mm after-filter. The impactor maintains high collection efficiency even at high particle concentrations, with minimal particle bounce and internal wall losses. It improves analytical sensitivity and minimises non-detectable samples. The collection substrate is chemically inert, and there is no impaction grease to contaminate the sample.

It is optimised at a 9 L/min flow rate, with low pressure drop for 24 hour sampling. Use it with the Leland Legacy sample pump at 9 L/min to ensure precise particle separation at the specified cut-points.

Which filters does the Sioutas impactor need?

  • Collection filter for the stages: 25 mm, 0.5 µm PTFE with laminated PTFE support (225-3708), required
  • After-filter: 37 mm, 2.0 µm PTFE with PMP support ring (225-1709), optional

Size-fractionated samples can be analysed gravimetrically, chemically and microscopically.

What is the Sioutas personal cascade impactor used for?

  • Occupational hygiene studies
  • Indoor Air Quality (IAQ) studies
  • Air pollution studies
  • Inhalation toxicology and epidemiological studies
  • Aerosol research
  • Pharmaceutical studies
  • Outdoor ambient sampling, which requires special provisions: see the Sioutas Cascade Impactor Instructions

For a single-stage impactor, see the Personal Modular Impactor (PMI). All size-selective samplers are listed under particle size-selective sampling.