Passive Samplers
10 models
Midget impingers, also called bubblers: small bottles that draw sampled air through a collection liquid to trap reactive gases and mists for wet-chemistry analysis. Offered in Pyrex glass — standard, fritted or spill-resistant midget bottles — and in unbreakable PFA, with holsters, holders and traps — from SKC.
Glass and PFA bubbler bottles drawing air through a liquid; 25 mL midget, fritted or spill-resistant.
25 mL · 170–200 µm frit
SKCAn impinger is a small bottle used with an air sample pump. Air is drawn through a collection liquid that traps reactive gases and mists for analysis. The SKC impingers include glass midget bottles of 25 mL. Their graduations are accurate to ±0.5 mL, which reduces the volumetric measurements needed in the field and in the laboratory.
Glass midget impingers come with a standard nozzle or a fritted nozzle with a 170–200 µm glass frit, which increases contact between the air sample and the liquid. The spill-resistant version has an outlet side arm that extends midway down the impinger with capacity for all the liquid in it. Both nozzle types are available in the spill-resistant design.
PFA impingers are unbreakable, inert to virtually all chemicals and rated to 260 °C, and they suit cryogenic applications. They come with a ¼ inch vertical port and a ¼ inch side port for horizontal connections, or with two vertical ¼ inch ports for close assembly of the sampling train. Holsters with a clip keep the impinger near the breathing zone, stainless steel holders suit area monitoring, and a trap stops impinger liquid from being drawn into the sample pump.
Our application engineers aim to reply within 3 business hours.