Gelatin Filters
Water-soluble gelatin filter, 3 µm nominal, pre-sterilised by gamma irradiation, for air sampling of bioaerosols; dissolves on agar.
The gelatin filter is a water-soluble filter for air sampling of bioaerosols, designed to maintain the viability of collected microorganisms. It has a nominal pore size of 3 µm and is supplied pre-sterilised by gamma irradiation in 25 mm (225-9551) and 37 mm (225-9552), packs of 50, without support pad. The high moisture content (46 to 49% residual dampness) helps maintain microorganism viability for sampling periods up to 30 minutes, and the filter dissolves easily when placed on agar, allowing a gentle transition from sample medium to growth medium. Being completely water soluble is also a prerequisite for virus sampling.
How well does a gelatin filter retain bacteria and phages?
Absolute retention rates, at inlet velocities of 0.25 m/s, 0.3 m/s (80 %RH) and 0.3 m/s (50 %RH) respectively:
- 99.9995% for Bacillus subtilis var. niger spores
- 99.94% for T3 phages (coliphages)
- 99.9% for T1 phages (coliphages)
Although the nominal pore size is 3 µm, a higher capture efficiency of sub-micron particles can be expected, because inertial impaction and diffusional interception remove particles much smaller than 3 µm.
How do I use a gelatin filter for air sampling?
Sample at 1 to 4 L/min for 30 minutes at typical indoor relative humidity, using a standard filter cassette, the Button Sampler or the IOM Sampler. The 25 mm filter 225-9551 is recommended for use with IOM and Button Samplers. For maximum culturability and superior collection of inhalable-size bioaerosols, combine 25 mm gelatin filters with the Button Sampler. Storage, shelf-life and conditions for use are in the Specifications tab.
Where are gelatin filters used?
The pre-sterilised filters suit monitoring in pharmaceutical plants and in areas where disinfectants or antibiotics are present. Typical applications are Indoor Air Quality (IAQ), outdoor environmental measurements, clean room monitoring, veterinary clinics, hospitals, pharmaceutical manufacturing and military. Compare other media in the air sampling filters and cassettes range.
| Parameter | Value |
|---|---|
| Material | Water-soluble gelatin |
| Pore Size | While having a nominal pore size of 3 µm, a higher capture efficiency of sub-micron particles can be expected due to the separations that occur on the surface and within the filter. It is through inertial impaction and diffusional interception that these filters can remove particles much smaller than 3 µm. |
| Thickness | Approximately 250 µm |
| Sterilisation | Presterilised by gamma radiation. Filters individually wrapped and packaged in units of five, each in a polyethylene bag. |
| Thermal Resistance | Maximum 60°C |
| Residual Dampness Content | 46 to 49% |
| Conditions for Use | Maximum room temperature is 30°C ; maximum relative humidity is 85%. |
| Sampling Time | 30 minutes at typical indoor relative humidity |
| Sampling Flow Rate | 1 to 4 L/min for filter sampling of biological contaminants using a standard filter cassette, Button Sampler, or the IOM |
| Storage | Gelatin filters may be stored in low humidity, ambient conditions; storage between 4 and 8°C is recommended. Caution: Do not store gelatin filters below 4°C. Condensation during thawing will dissolve filter. Avoid exposing filters to moisture, chemical vapors, and extreme temperatures. |
| Shelf-life | 3 years from date of manufacture |
| Analysis | Direct method or indirect method. Gelatin filters dissolve when placed on agar plates. |
| Diameter (mm) | Support Pad included | Notes | Pack Size | Part Number | |
|---|---|---|---|---|---|
| 25 | No | Water soluble | 50 | 225-9551 † * | |
| 37 | No | Water soluble | 50 | 225-9552 * |
Questions about this product?
Our application engineers aim to reply within 3 business hours.