GASTEC

Gastec Pyrotec Tubes

Gastec Pyrotec tubes, used with the Pyrolyzer for compounds which have to be thermally decomposed into substances to be measured.

Ask an engineer

Gastec Pyrotec tubes require the additional use of the portable Pyrotec Pyrolyzer (GST_840) accessory. Together they quickly, accurately and quantitatively measure fluorochlorocarbons, nitroparaffins, anesthetic gases, and other hard-to-detect compounds on-the-spot, as listed in the Specifications tab. The tubes belong to the Gastec colorimetric detector tubes.

How do Gastec Pyrotec tubes work with the pyrolyzer?

The sample enters the Pyrotec tube where it is thermally decomposed into substances that react chemically with the reagent to give the resultant colour change. The Pyrolyzer is not suitable for use in hazardous atmospheres.

Which Pyrotec tube for fluorochlorocarbons?

Tube choice depends on the compound and its range. For chlorodifluoromethane (R22), GST_51L reads 2.5 ppm – 135 ppm, GST_51 reads 25 ppm – 1000 ppm and GST_51H reads 0.1 % – 2.4 %. For acetonitrile, the nitro compounds tube GST_52 reads 3 ppm – 180 ppm. All other compounds and tubes are in the Specifications tab.

How are Pyrotec tubes supplied?

Tubes are supplied in packs of 10 unless noted otherwise (shown by + or #). Tubes marked + are twin tubes, sufficient for 5 tests: the first tube is a scrubber/oxidiser tube, the second tube provides the reading. Tubes marked # are twin tubes, sufficient for 4 tests.