title: "UMEx 300 passive sampler for ammonia — Analyt-MTC"
url: https://analyt-mtc.com/catalog/product/skc-umex-300-passive-sampler-ammonia/
lang: en
alternates:
  de: https://analyt-mtc.com/de/catalog/product/umex-300-passive-sampler-ammonia/
  en: https://analyt-mtc.com/catalog/product/skc-umex-300-passive-sampler-ammonia/
---

![UMEx 300 Passive Sampler Ammonia](https://analyt-mtc.com/illustrations/skc-umex-300-passive-sampler-ammonia.webp)

SKC

# UMEx 300 Passive Sampler Ammonia

UMEx 300 passive sampler, suitable for monitoring ammonia, one of the most commonly produced industrial chemicals, with chemistry similar to active OSHA Method ID-188 and NIOSH 6016

Documents

[SKC UMEx 300 Passive Sampler for Ammonia — Technical data sheet · Rev. 1](https://analyt-mtc.com/documents/skc-umex-300-spec-sheet-en.pdf) ([text version](https://analyt-mtc.com/documents/skc-umex-300-spec-sheet-en.md)) [SKC UMEx 300 Passive Sampler for Ammonia — Instructions for use · Rev. 2](https://analyt-mtc.com/documents/skc-umex-300-instructions-en.pdf) ([text version](https://analyt-mtc.com/documents/skc-umex-300-instructions-en.md))

The UMEx 300 passive sampler accurately and reliably samples workplace exposure to ammonia. Constructed of tough polypropylene, the single-use UMEx 300 contains a sulphuric acid treated tape for reliable collection of ammonia, with chemistry similar to active methods OSHA ID-188 and NIOSH 6016. It is used for personal or area sampling and is part of the [Passive Samplers](https://analyt-mtc.com/catalog/passive-samplers/) range.

-   Accurate and reliable
-   Highly sensitive and specific analysis method
-   Economical, low cost, single use sampler
-   Easy to use: no pump or training required
-   Sample medium and blank correction in one sampler
-   No particulate interference
-   Small and unobtrusive, with a simple-to-use “on/off” sliding cover
-   No glass or chemical liquids in the sampler

## How does the UMEx 300 passive sampler collect ammonia?

Samplers are provided in individual aluminiumised pouches that can be used to transport the sampler to a laboratory after sampling. For convenience and quality control assurance against contamination, each sampler incorporates a “blank correction” section in addition to the active sampling section, so there is no need to send extra samplers to the laboratory. Analysis is by ion chromatography (IC) with conductivity detection for the ammonium ion. The shelf-life date is printed on a label on the outside of each pouch for easy inventory management, and a clip attaches the sampler to a worker’s collar for personal sampling or to an appropriate location for area sampling.

## Which concentration range does the UMEx 300 cover?

The validated concentration range is 0.1 to 100 ppm. The sampling rate is 39.92 ml/min with a RSD of 11.4%, and the lower detection limit is 2.4 ppm for 15 minutes, 0.075 ppm for 8 hrs and 0.025 ppm for 24 hrs. The sampler has a limited shelf-life of 9 months from date of manufacture at <4°C and is designed for single use only. A pack holds 10 samplers (500-300).

## Which industries expose workers to ammonia?

Ammonia is one of the most commonly produced industrial chemicals. Workers can be exposed to ammonia in the following industries:

-   Agricultural (fertilisers)
-   Janitorial services (cleaners)
-   Food and beverage (refrigeration and fermentation)
-   Remediation (removal of contaminants)
-   Fuel
-   Manufacturing of plastics, explosives, pesticides, textiles and dyes

| Parameter | Value |
| --- | --- |
| Sampling Rate | 39.92 ml/min with a RSD (Relative Standard Deviation) of 11.4% |
| Validation Range | 0.1 to 100 ppm |
| Lower Detection Limits | 15 minutes: 2.4 ppm 8 hrs: 0.075 ppm 24 hrs: 0.025 ppm |
| Shelf-life | 9 months from date of manufacture at <4°C (39.2°F) |
| Analysis | Solvent extraction and ion chromatography with conductivity detection for the ammonium ion or visible absorption spectrometry |
| Accuracy | ± 23.4% |
| Storage | Before use: <4°C After use: Analyze within 3 weeks. Store at ambient temperature or at <4°C |
| Temperature Effects | No effect on sampling rate between 22 and 40°C |
| Humidity Effects | No effect from 20 to 80% relative humidity (RH) |
| Wind Effects | No effect from 5 to 100 cm/s |
| Interferences | None identified |
| Dimensions | 8.6 x 2.8 x 0.89 cm |
| Weight | 10.9 g |

| Description | Pack Size | Part Number |  |
| --- | --- | --- | --- |
| UMEx 300 Passive Sampler for Ammonia. / Designed for single use only. DO NOT reuse UME3 300 samplers. / Limited shelf-life; storage required at <4°C | 10 | **500-300** |  |
| Treated Tape, for QC (quality control) purposes only | 25 | **P20083** |  |

| Description | Part Number |  |  |
| --- | --- | --- | --- |
| Stand for Indoor Sampling | **690-302** |   ![UMEx 300 Passive Sampler Ammonia](https://analyt-mtc.com/illustrations/accessories/acc-690-302-3d.webp) |  |
| Shelter for Outdoor Sampling | **690-303** |   ![UMEx 300 Passive Sampler Ammonia](https://analyt-mtc.com/illustrations/accessories/acc-690-303-3d.webp) |  |

## Questions about this product?

Our application engineers aim to reply within 3 business hours.

[Ask an engineer](https://analyt-mtc.com/contact/?ask=1&about=UMEx%20300%20Passive%20Sampler%20Ammonia)