title: "UMEx 200 passive sampler for SO2/NO2 — Analyt-MTC"
url: https://analyt-mtc.com/catalog/product/skc-umex-200-passive-sampler-so2-no2/
lang: en
alternates:
  de: https://analyt-mtc.com/de/catalog/product/umex-200-passive-sampler-so2-no2/
  en: https://analyt-mtc.com/catalog/product/skc-umex-200-passive-sampler-so2-no2/
---

![UMEx 200 Passive Sampler SO2/NO2](https://analyt-mtc.com/illustrations/skc-umex-200-passive-sampler-so2-no2.webp)

SKC

# UMEx 200 Passive Sampler SO2/NO2

UMEx 200 passive sampler for occupational sampling and near-road/urban air monitoring, with the same chemistry as active OSHA Method ID-182

Documents

[SKC UMEx 200 Passive Sampler — Technical data sheet · Rev. 1](https://analyt-mtc.com/documents/skc-umex-200-spec-sheet-en.pdf) ([text version](https://analyt-mtc.com/documents/skc-umex-200-spec-sheet-en.md)) [SKC UMEx 200 Passive Sampler for Nitrogen Dioxide — Instructions for use · Rev. 2](https://analyt-mtc.com/documents/skc-umex-200-no2-instructions-en.pdf) ([text version](https://analyt-mtc.com/documents/skc-umex-200-no2-instructions-en.md)) [SKC UMEx 200 Passive Sampler for Sulphur Dioxide — Instructions for use · Rev. 2](https://analyt-mtc.com/documents/skc-umex-200-so2-instructions-en.pdf) ([text version](https://analyt-mtc.com/documents/skc-umex-200-so2-instructions-en.md))

The lightweight UMEx 200 passive sampler clips in the breathing zone for easy personal sampling of sulphur dioxide (SO2) and/or nitrogen dioxide (NO2). It is a single-use sampler with the same chemistry as active OSHA Method ID-182, and is part of the [Passive Samplers](https://analyt-mtc.com/catalog/passive-samplers/) range. The sample medium (tape treated with triethanolamine, TEA) and a “blank/correction” section of tape are contained in one sampler, saving time and money. Samplers are provided in individual aluminiumised pouches that can be used to transport the sampler to a laboratory; the label on the outside of each pouch contains the shelf-life date for easy inventory management.

-   Same chemistry as active OSHA Method ID-182
-   Ready to use
-   Accurate, reliable personal sampling
-   Lightweight sampler clips easily in the breathing zone
-   Economical, low cost, single use sampler
-   Safe: no glass or chemical liquids in the sampler
-   No pump or training required
-   Sample medium and blank/correction in one unit
-   3-week sample storage at ambient temperature

## Which exposures does the UMEx 200 passive sampler measure?

It accurately measures personal workplace exposure to sulphur dioxide and/or nitrogen dioxide from 15 minutes to 8 hours (not suitable for the new STEL of 0.25 ppm). The sampling rate is 15.2 ml/min for SO2 with a RSD of 16.5% and 17.3 ml/min for NO2 with a RSD of 11.4%.

## Is the UMEx 200 validated for ambient air?

SKC Research has validated the UMEx 200 for 24 hour ambient air sampling of sulphur dioxide and nitrogen dioxide to complement federal EPA reference methods. The validated concentration range for sulphur dioxide is 0.4 to 8 ppm (15 minutes to 24 hours); nitrogen dioxide is 0.05 to 8 ppm (15 minutes to 24 hours). Packs hold 10 (500-200) or 25 (500-200A) samplers, designed for single use only.

## Where are sulphur dioxide and nitrogen dioxide used?

Sulphur dioxide is used in:

-   Intermediate in the manufacture of sulphuric acid
-   Bleaching agent
-   Disinfectant
-   Fumigant
-   Solvent
-   Refrigerant
-   Food preservative
-   Reagent in the manufacture of magnesium, sodium sulphite and other chemicals

Nitrogen dioxide is used in:

-   Agricultural silos
-   Arc or gas welding, especially in confined space
-   Electroplating plants
-   Food and textile bleaching
-   Jewellery manufacturing
-   Nitric acid production
-   Nitrogen fertiliser production
-   Nitro-explosive production
-   Pickling plants

| Parameter | Value |
| --- | --- |
| Sampling Rate | SO2: 15.2 ml/min with a RSD (Relative Standard Deviation) of 16.5% NO2: 17.3 ml/min with a RSD of 11.4% |
| Validation Range | SO2: 0.4 to 8 ppm NO2 0.051 to 8.5 ppm |
| Lower Detection Limits | SO2: 15 minutes: 1.8 ppm 8 hrs: 52 ppb 24 hrs: 17.4 ppb NO2: 15 minutes: 0.2 ppm 8 hours: 0.3 ppb 24 hours: 2 ppb |
| Shelf-life | 18 months from date of manufacture at ambient temperature |
| Analysis | Solvent extraction and ion chromatography with conductivity detection |
| Accuracy | SO2: ± 30% NO2: ± 27% |
| Storage | Before use: ambient temperature After use: Analyze within 3 weeks. Can be stored at ambient temperature or at <4°C |
| Temperature Effects | SO2: No effect on sampling rate between 20 and 25°C NO2: No effect on sampling rate between 22 and 40°C |
| Humidity Effects | No effect from 20 to 80% relative humidity (RH) |
| Wind Effects | No effect from 5 to 100 cm/s |
| Interferences | None identified |
| Dimensions | 8.6 x 2.8 x 0.89 cm |
| Weight | 10.9 g |

| Description | Pack Size | Part Number |  |
| --- | --- | --- | --- |
| UMEx 200 Passive Sampler for Sulphur Dioxide and/or Nitrogen Dioxide, individually packaged in aluminised pouch. / UMEx 200 samplers are designed for single use only. / DO NOT reuse UMEx 200 samplers. / Limited shelf-life. | 10 | **500-200** |  |
| UMEx 200 Passive Sampler for Sulphur Dioxide and/or Nitrogen Dioxide, individually packaged in aluminised pouch. / UMEx 200 samplers are designed for single use only. / DO NOT reuse UMEx 200 samplers. / Limited shelf-life. | 25 | **500-200A** |  |
| Treated Tape, for QC (quality control) purposes only | 25 | **P20098** |  |

| Description | Part Number |  |  |
| --- | --- | --- | --- |
| Stand for Indoor Sampling | **690-302** |   ![UMEx 200 Passive Sampler SO2/NO2](https://analyt-mtc.com/illustrations/accessories/acc-690-302-3d.webp) |  |
| Shelter for Outdoor Sampling | **690-303** |   ![UMEx 200 Passive Sampler SO2/NO2](https://analyt-mtc.com/illustrations/accessories/acc-690-303-3d.webp) |  |

## Questions about this product?

Our application engineers aim to reply within 3 business hours.

[Ask an engineer](https://analyt-mtc.com/contact/?ask=1&about=UMEx%20200%20Passive%20Sampler%20SO2%2FNO2)