# Flite 4 Pump

SKC Flite 4 pump with lead-acid battery or AC mains: a portable and versatile pump for high volume air sampling, ideal for asbestos monitoring, with fully programmable run times, a backlit display for use in low light conditions and a robust steel casing

Documents

[SKC Flite4 High Volume Sample Pump — Operating manual · Rev. 2](https://analyt-mtc.com/documents/skc-flite4-manual-en.pdf) ([text version](https://analyt-mtc.com/documents/skc-flite4-manual-en.md)) [SKC Sampling Mast for Flite and MCS Series Pumps — Instructions for use · Rev. 2](https://analyt-mtc.com/documents/skc-sampling-mast-instructions-en.pdf) ([text version](https://analyt-mtc.com/documents/skc-sampling-mast-instructions-en.md))

The SKC Flite 4 pump is portable, versatile and fully programmable for high volume air sampling. The Flite4 air sample pump gives high performance sampling for dusts, aerosols, fibres, and microbes and is ideal for static asbestos sampling and long term background monitoring. Its flow range is 2–20 L/min, and it is listed under [high flow air sampling pumps](https://analyt-mtc.com/catalog/air-sampling-pumps-above-5-l-min/).

-   Portable and versatile pump for high volume air sampling
-   Ideal for asbestos monitoring
-   Fully programmable run times
-   Backlit display for use in dim/dark conditions
-   Robust steel casing with integral carrying handle
-   Small size: only 11 x 20 x 16 cm
-   Operates from battery or AC

## What is the Flite 4 pump used for?

Applications:

-   Area and environmental monitoring
-   Biosampling
-   Indoor air studies
-   Asbestos monitoring
-   Long-term background monitoring
-   Stack sampling
-   Preferred Vacuum Test (PVT) for ductwork, per BS EN 15780:2011 or BESA TR19

## How is the Flite 4 pump powered, and how long does it run?

The pump runs from a 7 Ah rechargeable lead-acid battery pack, an external high capacity lead-acid battery (a 12 Ah battery, P901107, requires cable 901-212) or AC mains. The high capacity (7Ah) battery model gives longer run times, over 12 hours at 8 L/min, and charging takes 10 hours for the 7 Ah battery. The LCD shows low battery and fault indication, and a flow fault is displayed if flow is restricted.

## What does the Flite 4 pump kit include?

The Flite4 sample pump kit (901-4011K) includes the pump with 7 Ah battery, a single charger and 1 m length Tygon tubing. The pump is also available without battery (901-4011). It needs 1/4″ ID tubing. The Specifications tab lists the accessories, such as sampling masts and the chek-mate flowmeter.

| Parameter | Value |
| --- | --- |
| Flow System | – Flow range: 2,000 to 20,000 ml/min – Flow fault displayed if flow is restricted |
| Power Supply | – 7 Ah Rechargeable Lead-Acid battery pack, external high capacity Lead-Acid battery or AC Mains – Low battery indication on LCD – Charging time: 10 hours (7 Ah battery) |
| Display | – Backit LCD – Displays elapsed run time, timed run time, delayed start run depending on the programme – Low battery and fault indication |
| Temperature Range | \-5 to 50°C |
| Case | Steel casing with integral carrying handle |
| Tubing Required | 1/4″ ID |
| Dimensions | 11.2 x 20.3 x 16.3 cm |
| Weight | – 2060 g (no battery) – 4260 g (fitted with 7 Ah battery) |

| Description | Part Number |  |
| --- | --- | --- |
| Flite4 Sample Pump without battery | **901-4011** |  |

| Description | Part Number |  |
| --- | --- | --- |
| **Flite4 Sample Pump Kit includes:** / Flite4 sample pump with 7 Ah battery, single charger and 1 m length Tygon tubing | **901-4011K** |  |

| Description | Part Number |  |  |
| --- | --- | --- | --- |
| Single Charger 100-240 V with UK/EU/US/AUS mains plugs | **901-410** |   ![Flite 4 Pump](https://analyt-mtc.com/illustrations/accessories/acc-901-410-front.webp) |  |
| chek-mate flowmeter, 5.0-30.0 l/min (recommended) | **375-50300** |   ![Flite 4 Pump](https://analyt-mtc.com/illustrations/accessories/acc-chek-mate-in-hand-500x500.webp) |  |
| Pulsation Dampener, for use with chek-mate 375-50300 | **375-150** |  |  |
| Rigid Aluminium Sampling Mast – two piece, 1 m high | **901-213** |   ![Flite 4 Pump](https://analyt-mtc.com/illustrations/accessories/acc-901-213-3d.webp) |  |
| Extension Piece for Rigid Aluminium Sampling Mast | **901-213A** |  |  |
| Rigid Aluminium Sampling Mast – four piece, 1 m high | **901-214** |  |  |
| Hosetail Adaptor for 901-213 and 901-214 Sampling Masts | **901-215** |  |  |
| Replacement Lead-Acid Battery (7 Ah) | **P901402** |   ![Flite 4 Pump](https://analyt-mtc.com/illustrations/accessories/acc-p901402-3d.webp) |  |
| External High Capacity 12 Ah Lead-Acid Battery (requires 901-212 cable) | **P901107** |   ![Flite 4 Pump](https://analyt-mtc.com/illustrations/accessories/acc-p901107-3d.webp) |  |
| Flite4 to External Battery (P901107) Cable | **901-212** |   ![Flite 4 Pump](https://analyt-mtc.com/illustrations/accessories/acc-901-212-front.webp) |  |
| Flite Series 100-240V AC Adaptor with UK/EU/US/AUS mains plugs | **901-411** |   ![Flite 4 Pump](https://analyt-mtc.com/illustrations/accessories/acc-901-411-front.webp) |  |

## Questions about this product?

Our application engineers aim to reply within 3 business hours.

[Ask an engineer](https://analyt-mtc.com/contact/?ask=1&about=Flite%204%20Pump)