title: "HAZ-DUST IV personal particulate monitor — Analyt-MTC"
url: https://analyt-mtc.com/catalog/product/haz-dust-iv-particulate-monitor/
lang: en
alternates:
  de: https://analyt-mtc.com/de/catalog/product/haz-dust-iv-particulate-monitor-2/
  en: https://analyt-mtc.com/catalog/product/haz-dust-iv-particulate-monitor/
---

![HAZ-DUST IV Particulate Monitor, 3D view](https://analyt-mtc.com/media/hazdust-IV_3D.png)

1 / 3

SKC

# HAZ-DUST IV Particulate Monitor

The HAZ-DUST IV particulate monitor is a personal real-time aerosol monitor providing immediate breathing zone measurements of aerosol and dust for industrial hygiene and environmental air investigations

Documents

[SKC HAZ-DUST IV Particulate Monitor — Operating manual · Rev. 2](https://analyt-mtc.com/documents/skc-haz-dust-iv-manual-en.pdf) ([text version](https://analyt-mtc.com/documents/skc-haz-dust-iv-manual-en.md)) [SKC HAZ-DUST IV Calibration Span Reference — Instructions for use · Rev. 2](https://analyt-mtc.com/documents/skc-haz-dust-calibration-standard-instructions-en.pdf) ([text version](https://analyt-mtc.com/documents/skc-haz-dust-calibration-standard-instructions-en.md)) [SKC Adapted IOM Inlet for HAZ-DUST — Instructions for use · Rev. 2](https://analyt-mtc.com/documents/skc-haz-dust-iom-adapter-instructions-en.pdf) ([text version](https://analyt-mtc.com/documents/skc-haz-dust-iom-adapter-instructions-en.md)) [HAZ-DUST-IV-QS-manual](https://analyt-mtc.com/media/HAZ-DUST-IV-QS-manual.pdf)

The HAZ-DUST IV particulate monitor is a personal real-time aerosol monitor for occupational hygiene and environmental air investigations. It displays immediate breathing zone measurements and comprehensive graph reporting of aerosol and dust, with a sensing range of 0.01 to 200 mg/m3 and a sample flow of 1.0 to 3.3 L/min (adjustable). It belongs to the [particulate monitors](https://analyt-mtc.com/catalog/particulate-monitors/) range.

-   Easy operation and instantaneous, continuous display of TWA, STEL, Minimum, Maximum concentration in mg/m3
-   Sensor mounts directly in the breathing zone for a personal sample
-   Small and lightweight with integral belt clip
-   Operates for over 8 hours using rechargeable NiMH battery

## Which sampling head suits which particle fraction?

Interchangeable sampling heads for Inhalable, Thoracic and Respirable measurements let you target the fraction you need. The monitor kit includes a thoracic sampling head.

An external switch selects between respirable calibration to NIOSH 0600 and 2.75 L/min flow rate (GS-3 Cyclone) or inhalable calibration to NIOSH 0500 and 2 L/min flow rate (IOM Sampler).

## How do I take a gravimetric back-up sample?

An in-line 37 mm cassette (or optional 25 mm cassette using adaptor) collects a concurrent filter sample for gravimetric or chemical analysis. For a passive alternative to this monitor, see the [passive particulate monitor](https://analyt-mtc.com/catalog/product/split2-particulate-monitor/).

## What does the HAZ-DUST IV record?

Automatic data logging stores readings, and the easy to use [DustComm Pro Software](https://www.skcltd.com/products2/particulate-monitors/dustcomm-pro-software.html) handles data downloading, trend analysis and comprehensive reporting of concentration changes over time.

| Parameter | Value |
| --- | --- |
| Accuracy | ±10% to filter gravimetric SAE fine test dust |
| Sensing Range | 0.01 to 200 mg/m3 |
| Display Resolution | 0.01 mg/m3 |
| Particle Size Range | 0.1 to 100 µm |
| Measurement Precision | ±0.02 mg/m3 |
| Calibration | NIOSH Method 0600 gravimetric reference – NIST-traceable SAE fine test dust |
| Displayed Particle Concentration | Instantaneous, TWA, STEL, Minimum, Maximum |
| Sample Flow Rate | 1.0 to 3.3 L/min (adjustable) – externally switchable 2 L/min or 2.75 L/min |
| Power | Rechargeable NIMH battery |
| Operating Time | ≥8 hours |
| Charging Time | 10 to 12 hours |
| Digital Output | RS-232 |
| Data Storage | 21,500 data points |
| Temperature | Operating: 0 to +50°C Storage: -20 to +70°C |
| Dimensions | Case: 14 x 8.3 x 7 cm Sensor: 4.4 x 3.8 |
| Weight | 680 g |

| Description | Part Number |  |
| --- | --- | --- |
| HAZ-DUST IV Monitor Kit, includes: / Monitor, Thoracic sampling head, in-line 37 mm cassette, zeroing filter for inhalable dust, carry case, computer cable, AC Charger 110-240 V, HAZ-DUST Media USB with instruction manual and DustComm Pro Software / (requires calibration with equipment sold separately, see accessories below) | **770-4004** |  |

| Description | Part Number |  |  |
| --- | --- | --- | --- |
| IOM Sampling Head with 25 mm cassette | **225-70A** |   ![HAZ-DUST IV Particulate Monitor](https://analyt-mtc.com/illustrations/accessories/acc-iom-sampler-3d.webp) |  |
| IOM Adaptor, for mounting IOM sampling head on HAZ-DUST IV | **770-4204** |   ![HAZ-DUST IV Particulate Monitor](https://analyt-mtc.com/illustrations/accessories/acc-770-4204-3d.webp) |  |
| Inhalable Impactor and Adaptor | **770-115A** |   ![HAZ-DUST IV Particulate Monitor](https://analyt-mtc.com/illustrations/accessories/acc-770-115a-3d.webp) |  |
| Thoracic Sampling Head, mounts on inlet of HAZ-DUST IV sensor | **770-4107** |   ![HAZ-DUST IV Particulate Monitor](https://analyt-mtc.com/media/770-4107_3D-300x300.png) |  |
| Respirable Impactor, GS-3 Cyclone, mounts on HAZ-DUST IV sensor | **225-103** |   ![HAZ-DUST IV Particulate Monitor](https://analyt-mtc.com/illustrations/gs-3-respirable-dust-cyclone.webp) |  |
| Adaptor for Respirable Impactor, required when using GS-3 Cyclone | **770-313** |   ![HAZ-DUST IV Particulate Monitor](https://analyt-mtc.com/illustrations/accessories/acc-770-313-3d.webp) |  |
| Calibration Standard, for verifying span and optical sensor performance | **770-140** |   ![HAZ-DUST IV Particulate Monitor](https://analyt-mtc.com/media/770-140_3D-300x300.png) |  |
| Calidaptor, adaptor for calibrating the Inhalable IOM Sampling Head | **391-01** |   ![HAZ-DUST IV Particulate Monitor](https://analyt-mtc.com/illustrations/accessories/acc-391-01-3d-1.webp) |  |
| Recommended chek-mate flowmeter, 500-5000 ml/min, for calibrating the Inhalable Sampling Head | **375-0550** |   ![HAZ-DUST IV Particulate Monitor](https://analyt-mtc.com/illustrations/accessories/acc-chek-mate-in-hand-500x500.webp) |  |
| Calibration Chamber, for calibrating and setting pump airflow when using GS-3 Cyclone | **225-112** |   ![HAZ-DUST IV Particulate Monitor](https://analyt-mtc.com/illustrations/accessories/acc-225-111-3d.webp) |  |
| Zeroing Accessory, for use with Inhalable Impactor | **770-4202A** |   ![HAZ-DUST IV Particulate Monitor](https://analyt-mtc.com/media/770-4202A_3D-300x300.png) |  |
| Zeroing Accessory, for use with Thoracic Impactors | **770-4102** |   ![HAZ-DUST IV Particulate Monitor](https://analyt-mtc.com/media/770-4102_3D-300x300.png) |  |
| Cleaning Kit, includes bulb, swabs and lens tissue | **770-305B** |  |  |
| AC Charger 110-240 V | **770-319** |  |  |
| Replacement NiMH Battery Pack | **770-4105** |   ![HAZ-DUST IV Particulate Monitor](https://analyt-mtc.com/media/770-4105_3D-300x300.png) |  |
| Protective Pouch | **770-325** |  |  |
| DustComm Pro Software on USB Flash Drive | **770-225** |  |  |

## Questions about this product?

Our application engineers aim to reply within 3 business hours.

[Ask an engineer](https://analyt-mtc.com/contact/?ask=1&about=HAZ-DUST%20IV%20Particulate%20Monitor)