Aalborg

# DFC Mass Flow Controller

DFC mass flow controller: programmable thermal mass flow controller with digital signal processing, RS-232/RS485 interface and flow ranges up to 100 sL/min N2.

Documents

[Aalborg DFC Mass Flow Controllers — Technical data sheet · Rev. 2](https://analyt-mtc.com/documents/dfc-spec-sheet-en.pdf) ([text version](https://analyt-mtc.com/documents/dfc-spec-sheet-en.md)) [Aalborg DFC Digital Mass Flow Controller — Operating manual · Rev. 2](https://analyt-mtc.com/documents/dfc-manual-en.pdf) ([text version](https://analyt-mtc.com/documents/dfc-manual-en.md))

The DFC mass flow controller from Aalborg is a compact digital controller for the flow rate of gases, available as DFC 26/36/46. Using thermal mass flow metering, it offers an accuracy of ± 1.0% full scale at calibration temperature and pressure, a repeatability of ± 0.15% of full scale and a digital interface. It is part of the [thermal mass flow controllers](https://analyt-mtc.com/catalog/thermal-mass-flow-controllers/) range and works with dry gases only.

DFC mass flow controllers can be configured for flow ranges from 10 smL/min to 100 sL/min N2. Maximum gas pressure is 70 bar(g). For a low priced alternative in the same range, compare the [GFC](https://analyt-mtc.com/catalog/product/gfc/).

## How do I control and read a DFC?

These controllers can be accessed remotely using the RS-232/RS485 interface. The mass flow can also be accessed by the analog output either in 0-5Vdc/0-10Vdc or 4-20mA. Totalizer, flow and pressure alarms can also be set by the user.

## Which DFC model and fittings do I need?

The meters can be connected to the process with 1/8″, 1/4” or 3/8” compression fittings, depending on the model:

-   DFC 26: 1/8″ compression fittings
-   DFC 26/36: 1/4″ compression fittings
-   DFC 26/36/46: 3/8″ compression fittings

Optional fittings are available. They can be used as benchtop units or mounted with screws in the base.

| Parameter | Value |
| --- | --- |
| **Accuracy** | ± 1.0% full scale at calibration temperature and pressure. |
|  | Performed at standard conditions (1013 hPa) & (21.11°C). |
| **Repeatability** | ± 0.15% of full scale. |
| **Response Time** | 1.0 to 2.0 second to within ±2% of flow rate, over 20% to 100% full scale |
| **Gas Temp. & ambient Temp.** | Gas temp (0°C to 50°C). / Ambient temp (-10°C to 50°C). / Dry gases only. |
| **Output signals** | Linear 0-5 Vdc (2000 ohms min load impedance); 0-10Vdc (4000 ohms min impedance); / 4-20 mA optional (0-500 ohms\\ loop resistance). Maximum noise 20mV peak to peak. |
| **Gas pressure** | Max. 70 bar(g) |
| **Materials** | 316 stainless steel, 416 stainless steel, Viton® O-rings. Optional |
| **Connections** | DFC 26: 1/8″ compression fittings. / DFC 26/36: 1/4″ compression fittings. / DFC 26/36/46: 3/8″ compression fittings. |