title: "SKC BioStage impactor, 28.3 L/min, D50 0.6 µm — Analyt-MTC"
url: https://analyt-mtc.com/catalog/product/biostage/
lang: en
alternates:
  de: https://analyt-mtc.com/de/catalog/product/biostage-2/
  en: https://analyt-mtc.com/catalog/product/biostage/
---

![BioStage](https://analyt-mtc.com/illustrations/biostage.webp)

SKC

# BioStage

The SKC BioStage single-stage cascade impactor collects viable bacteria and fungi from indoor and outdoor air onto 90 to 100 mm agar plates at 28.3 L/min, and meets NIOSH Method 0800 and 0801.

Documents

[SKC BioStage Viable Cascade Impactor — Technical data sheet · Rev. 1](https://analyt-mtc.com/documents/skc-biostage-spec-sheet-en.pdf) ([text version](https://analyt-mtc.com/documents/skc-biostage-spec-sheet-en.md)) [SKC BioStage Impactor — Operating manual · Rev. 2](https://analyt-mtc.com/documents/skc-biostage-manual-en.pdf) ([text version](https://analyt-mtc.com/documents/skc-biostage-manual-en.md))

The SKC BioStage is a single-stage cascade impactor for sampling viable microorganisms, including bacteria and fungi, in indoor and outdoor air. It operates on the principle of inertial impaction and meets NIOSH Method 0800 and 0801 requirements and ACGIH recommendations for bioaerosol sampling. Its performance is equivalent to Andersen N-6 and Aerotech A6. For choosing, the specs that matter are the 28.3 L/min flow rate, the median cut-point (D50) of 0.6 µm and the 90 to 100 mm agar plate; the rest is on the Specifications tab.

## How does the BioStage single-stage impactor work?

A pump draws air through the jet stage, which has 400 holes (0.25 mm hole diameter), and particles impact onto the agar collection medium. Collected organisms remain intact and viable for colony culture. The precision-tooled aluminium parts are corrosion resistant and autoclavable, while the BUNA-N O-rings are not. The SureLock positive seal holds the components together and ensures sample integrity. It is easy to use, with easy setup and calibration, and operates on a proven principle.

## Which pump and agar plates does the BioStage need?

Use a high flow pump such as the [BioLite+ Pump](https://analyt-mtc.com/catalog/product/skc-biolite-pump/); the tubing adaptor 225-6202 connects it to the BioStage, which takes 1/4 inch ID tubing. Suggested media:

-   Bacteria: Tryptic Soy Agar (TSA) or Blood Agar Plates (BAP)
-   Viable fungi: Potato Dextrose Agar (PDA), Malt Extract Agar (MEA), Dichloran Glycerol 18 Agar (DG-18) or Corn Meal Agar (CMA)

Accessories: the calibration adaptor P33100, which attaches to the BioStage inlet, and the 1.5m telescoping tripod stand 228-506 for area sampling at breathing zone height. To collect into liquid instead of onto agar, see the [BioSampler](https://analyt-mtc.com/catalog/product/biosampler/). All viable air samplers are listed under [bioaerosol samplers](https://analyt-mtc.com/catalog/bioaerosol-samplers/).

## Where is the BioStage used?

-   Indoor Air Quality (IAQ) studies
-   Filter and cleanroom efficiency studies
-   Pharmaceutical production, cosmetic manufacturing and food processing areas
-   Brewery fermentation and grain processing and transportation
-   Animal care laboratories, hospital environments and sewage treatment plants
-   Bio-risk response

| Parameter | Value |
| --- | --- |
| Flow Rate | 28.3 L/min |
| Material | Inlet cone and base plate: Precision-tooled autoclavable aluminum O-rings: Duro 50, BUNA-N (not autoclavable ) |
| Jet Classification Stage | 400 holes (0.25 mm hole diameter) |
| Median Cut-Point (D50) | 0.6 μm |
| Sample Media | 90 to 100 mm agar plates \* |
| Suggested Media | **For bacteria:** Tryptic Soy Agar (TSA) Blood Agar Plates (BAP) **For fungi (viable):** Potato Dextrose Agar (PDA) Malt Extract Agar (MEA) Dichloran Glycerol 18 Agar (DG-18) Corn Meal Agar (CMA) |
| Analysis | Colony culture ‡ |
| Tubing | 1/4 inch ID |

| Description | Part Number |  |  |
| --- | --- | --- | --- |
| Standard BioStage single-stage cascade impactor | **225-9611** |   ![BioStage](https://analyt-mtc.com/illustrations/biostage.webp) |  |
| Calibration Adaptor for BioStage, attaches to the BioStage inlet | **P33100** |   ![BioStage](https://analyt-mtc.com/illustrations/accessories/acc-p33100-3d.webp) |  |
| Tubing Adaptor for BioLite+ Pump to BioStage | **225-6202** |   ![BioStage](https://analyt-mtc.com/illustrations/accessories/acc-225-6202-3d-1.webp) |  |
| Tripod Stand, 1.5m telescoping, for area sampling at breathing zone height | **228-506** |   ![BioStage](https://analyt-mtc.com/illustrations/accessories/acc-228-506-3d-1.webp) |  |

## Questions about this product?

Our application engineers aim to reply within 3 business hours.

[Ask an engineer](https://analyt-mtc.com/contact/?ask=1&about=BioStage)